1.4.2 Contraindications
DO NOT USE:
■
with life-sustaining critical application drugs, or drugs with short half-life;
■
for enteral administration;
■
if parenteral nutrition is contraindicated;
■
for infusion of insulin;
■
with neonates.
1.4.3 Intended user profile
The pump can only be used by qualified and trained healthcare professionals, trained
patients or trained patient relatives.
Patient and patient relatives must only use the pump under the responsibility of a qualified
and trained healthcare professional.
It is recommended that user attend an at least 60 minutes initial training session. It is
recommended that users attend a refresher training session every year. For training, contact
your Fresenius Kabi sales representative.
INFORMATION
Product specialists are in charge of configuration and maintenance of pumps
according to the technical manual.
1.4.4 Intended patient population
The pump is intended for use on adults and children. It cannot be used on neonates.
The pump is intended for use in accordance with healthcare facilities and healthcare providers
protocols. All pump settings must be made according to the medical prescription.
1.4.5 Use environment
WARNING
■
Keep the pump, sets and cables away from unsupervised children.
■
Keep the pump, sets and cables away from pest, animals and pets.
■
Keep away from heat sources, dust, fluff, direct and prolonged light
exposure.
■
The pump should be used under specified operational conditions listed
below to ensure pump performance.
Ambix nova is an ambulatory infusion system intended to be used in a home environment
(homecare and nursing home) and in ambulatory care environment.
The pump should be used in the following operational conditions to ensure proper
performance:
■
Operating temperature range: 5°C to 40°C
■
Operating humidity range: 20% ~ 85%, no condensation
9