beep sounds (if sound level is low, melody is playing on low, if sound level is medium,
melody is on medium, if sound level is high, melody is on high).
NOTE:
When display switches from switch on screen to main screen, the main screen
displays the last validated infusion settings.
5.3 Preparing the set
In order to protect user health from infection, please follow aseptic handling procedures for
container, set and infusion tube.
Only Fresenius Kabi infusion sets can guarantee pump reliability (
on
page 102)
WARNING
■
Do not use incompatible sets, it can lead to patient harm.
■
Check the infusion set and patient connection integrity before use to protect
the patient from underdose, trauma and infection.
■
Change the set after 24 hours of use to avoid therapeutic issues.
5.3.1 Description of the pinch clamp
Pinch clamp is closed
Pinch clamp is open
INFORMATION
Patient must not be connected to the set when the pinch clamp is open.
5.4 Loading the set
To connect / disconnect or change the container and set, please refer to the Instructions
for Use of the administration set.
When using the pump on a patient requiring special attention, ensure that a backup gravity
set is available for immediate use specially in homecare environment.
When the set is load in the pump, it must not be connected to the patient.
27