5.5 Priming the set with the pump (recommended)
WARNING
■
Patient must not be connected to the set when priming is performed.
■
Check the air absence in the infusion set at the end of priming.
INFORMATION
■
To proceed to infusion set priming, fill drip chamber 2/3 full by pressing
gently if applicable.
■
Automatic and semi-automatic priming fill the infusion set at a rate of 600
mL/h and are stopped after 24 mL ± 10%.
■
During priming, the air in line alarm is disabled.
5.5.1 Access priming interface
Priming interface is only accessible when pump is on infusion program interface (the following
screen illustrates these principles).
Press prime key