3.
Check the administration set to make sure there is no air in the line.
4.
If necessary, press and hold
(
) key to perform semi-automatic priming (see
on page 31) until no air can be found in the line.
5.
Press
(
) key to confirm. Pump automatically goes back to the infusion
programming screen as illustrated below.
Note
: Automatic priming can be stopped at any time by pressing the
key. Once it is
stopped, it is possible to continue priming by using
on page 31.
5.5.3 Semi-automatic priming
Check the administration set to make sure it is not connected to the patient.
1.
Under priming interface (see
on page 29 to access),
1
press
and
Hold
(
) key to prime and
2
release it when priming is complete.
31