1.
Press
or
on the keypad to select the parameter needs to be adjusted and press
to adjust the value of the selected parameter (the selected parameter appears
in a dark blue background).
2.
Repeat step 1 until all parameters are set.
Incompatible values
When a new infusion is programmed with incompatible values:
■
The selected parameter cannot be set to an incompatible value.
■
The linked parameter to the parameter adjusted is out of setting range, the background
changes to orange until the setting is compatible.
When infusion is stopped, and current parameters are changed:
■
If the parameter adjusted is out of setting range, the selected and the linked values
background flash in black.
Once all parameters are set, go to
on page 33.
INFORMATION
■
A longer keypress provides faster scrolling.
■
The flow rate of administration must be adapted individually to the patient.
Regular checks are required.
■
The selected parameter appears in a dark blue background and the linked
parameter appears in a light blue.
■
The linked parameter is calculated automatically when the other parameters
are programed.
■
For all modes, it is possible to change the infusion parameters when infusion
is stopped.
■
After switch on the pump, the pump stays on the last validated infusion
programming screen.
5.7 Start infusion
1.
Check the administration set integrity.
2.
Check that no air remains in the administration set.
33