1.
Under advanced settings menu (see
on page 60 to access),
select reset manufacturing menu
and press
(
) to enter.
2.
Press
to activate reset manufacturing settings.
3.
Press
(
) to confirm.
Once reset is confirmed, the reset symbol displays and rotates for 3 seconds, the pump
returns to the menu screen.
INFORMATION
When reset manufacturing settings is done:
■
Pump settings revert back to factory settings (except date and time settings).
■
The records in the history menu are erased and back to factory settings:
including infusion history, alarm history and cumulated counter.
6.7 History
Under main menu screen (see
on page 40), select the history menu
and
press
(
) to access the history menu.