8 Quick check protocol
The following checks allow user to confirm device behaviour according to these instructions
for use. Fresenius Kabi recommends to perform these tests before connecting Ambix nova
to patient.
WARNING
Do not use the device in case of check test failure and contact the appropriate
department or Fresenius Kabi sales representative for additional verification.
Yes
Action
Before use
1 - Check if the Ambix nova pump, holder and power cable are not damaged in any
way
2 - Check the general status of the display
3 - Install the Ambix nova pump on the holder
4 - Connect holder to the mains
5 - Switch on the pump
6 - Check the autotest sequence
(LCD display intact, speaker, LED and the backlight). Do not use with a damaged
screen.
7 - Check the mains LED lights up
8 - Remove the Ambix nova pump from its holder and check the battery symbol on the
display
9 - Install the Ambix nova pump on the holder
10 - Check that the pump and its holder are securely attached or positioned
11 - Check that all menu settings are adapted to the next patient
12 - Connect a set to a filled container, install the set in the pump and close the door
13 - Prime the set
14 - Set pump at prescribed parameters
15 - Start infusion
16 - Check the infusion information (droplet animation)
17 - Check that pumping is effective
After use
1 - Check if pump, holder and power cable are not damaged in any way
2 - Clean the pump, holder and power cable
77