11.1.8 Empty bag / Air in Line detection time at different flow rates
The table below illustrates the single bubble air detection time at its maximum selectable
volume of 0.75 mL.
Empty bag / Air in Line detection time
Air volume of single bubble = 0.75 mL
Flow rate
46 minutes maximum
1 mL/h
2 minutes maximum
25 mL/h
35 seconds maximum
100 mL/h
7 seconds maximum
600 mL/h
INFORMATION
The detection time of cumulated air bubbles is customized by the user. To set
the air detection volume and the detection time of cumulated air bubbles, see
Occlusion pressure / Air detection
on page 61.
11.2 Technical characteristics
11.2.1 Operation mode
The Ambix nova pump is a reusable device. The pump ensures fluid delivery in 4 kinds of
infusion mode (for
on page 44), using pumping and clamping fingers to push
the liquid to the patient.
11.2.2 Power supply specifications
The power cable must be connected directly to the mains power socket.
AC input voltage: 100-240 Vac
AC input frequency: 50/60 Hz
AC input current: 110 mA-205 mA
Smart Holder input
9 Vdc ± 5 % / 9 W (maximum load)
Smart Holder output
Approx. 2 m
Power cable length
11.2.3 Battery specifications
7.2 V Lithium Ion rechargeable battery / 16.2 W
Characteristics
Approximately 0.095 kg
Weight
6 h with pump off
Maximum charging time
11.2.4 Power consumption
Consumption of the pump in standard operating conditions: maximum 9 W.
90