12 Transport, storage and recycling conditions
12.1 Storage and transport conditions
WARNING
The pump must be stored and transported under
specified conditions
listed
below to ensure pump performance and to avoid malfunction.
Observe the following conditions for storage and transport:
■
Temperature range:
- 20°C to + 45°C
■
Humidity range:
10% to 85%, no condensation
■
Pressure range:
700 hPa to 1060 hPa
■
Altitude range:
less than 3000 m
12.1.1 Storage
Please make sure the pump is stored in an appropriate manner to avoid pump malfunctioning.
INFORMATION
■
The storage area must be clean, organized and compliant with the storing
conditions mentioned above.
■
The Ambix nova pump must be handled with care during storage.
12.1.1.1 Prepare the device for storage
WARNING
Remove the battery and store it as per storage conditions above in case of a
prolonged storage period
.
To prepare the device before storage, proceed as specified below:
1.
Be sure the pump is not being used on a patient.
2.
Switch pump OFF and remove installed set (see
Removing/Changing the set from the
on page 38).
3.
Disconnect pump power cable (see
on page 22).
4.
Remove the pump and its holder from pole or rails (see
on page 22).
5.
Clean the pump (see
on page 74).
6.
Handle the pump with care and store it in a compliant area.
12.1.1.2 Install the device after storage
INFORMATION
■
If the battery has been removed for storage, please contact your biomedical
department to replace the battery into the device prior to using the pump.