46
FONIX FP35 Quick Reference Guide
5.1 Target Creation
In order to create a target for use in the Coupler Target or Coupler EarSim screen, you
must enter the patient’s threshold values in the Audiogram Entry screen, just as you
would if you were creating a real-ear target.
To create a target in the Target Coupler screen:
1. Enter the Audiogram Entry screen by pressing [F2] from the Opening screen and
using the [NEXT] and [BACK] keys if necessary.
2. Follow the instructions found in Section 4.3 to enter the patient’s audiogram and
to create a real-ear target.
3. Follow the instructions found in Section 4.3.4, if desired, to measure the
patient’s RECD.
4. Press [EXIT] to return to the Opening screen.
5. Press [F3] to enter one of the coupler multicurve screens. If the screen is labeled
COUPLER MULTICURVE (the label is found on top of the large graph), press
[NEXT] to enter the Target Coupler screen.
6. Press [MENU] and [BACK] to enter the Target menu. Set the AID TYPE, FIT
RULE, and AGE in this menu using the arrow keys. If you are creating an NAL-
NL1 coupler target, you will also have to set COMPRESSION, CHANNELS, AID
LIMITING, FIT TYPE, TUBING, and VENT. See Section 4.3.3 for details.
7. Press [EXIT] to return to the measurement screen. You are now ready to perform
the coupler measurements.
5.2 Coupler Target
Press [NEXT] from the Coupler Multicurve screen or [BACK] from the Coupler
EarSim screen to enter the Coupler Target screen.
In the Coupler Target screen, the real-ear target is converted into a coupler target
displayed in either dB GAIN or dB SPL. Coupler measurement can be performed and
compared directly to the target. The only major difference between this screen and
the basic Coupler Multicurve screen (described in Chapter 2), is the addiition of the
coupler target. See Figure 5.2.
To get the most accurate real-ear to coupler target conversions, measure the patient’s
unaided response in one of the real-ear measurement screens (Section 4.4.1) and
the RECD in the Audiogram Entry screen (Section 4.3.4). When these measurements
are not taken, average values, corrected for the age of the patient, are substituted
automatically.
See Section 2.2 for instructions on how to perform measurements in the Coupler
Target screen—the operation of the Coupler Target screen is identical to the operation
of the Coupler Multicurve screen.