5: 2-cc Targets
47
Figure 5.2—Coupler Target screen.
5.3 Coupler EarSim
Press [NEXT] from the Coupler Target screen or [BACK] from the Coupler Multicurve
screen to enter the Coupler EarSim screen.
In the Coupler EarSim screen, the actual real-ear target prescribed for the patient
is displayed in either dB IG (insertion gain) or dB SPL. When the Coupler EarSim
display is set to dB SPL, the patient’s threshold and uncomfortable values are
also displayed. Any coupler measurement made in this screen is converted into
simulated real-ear measurements that are directly comparable to the displayed real-
ear target.
To get the most accurate real-ear to coupler target conversions, measure the patient’s
unaided response in one of the real-ear measurement screens (Section 4.4.1). The
measured REUR will be used to convert the coupler measurements into simulated
real-ear measurements. See Section 4.4.1 for instructions on performing the REUR
measurement.
In the SPL display (Figure 5.3), the patient’s audiometric data (HTL and UCL values)
are converted from dB HL to dB SPL and displayed on the screen. To get the most
accurage HL to SPL conversion (and the most accurage real-ear target), measure the
patient’s RECD in the Audiogram Entry screen. See Section 4.3.4 for details.
When the REUR and RECD are not measured, the FP35 substitutes average data
to calculate the real-ear target and the coupler measurement conversions. These
averages are sensitive to the selected age of the patient.