11
8/2021
The FSN wireless imaging system is intended to be used as a pair (transmitter and receiver)
to redundantly display video where needed. The wireless device is a non-sterile, reusable
device, not intended for use in the sterile fi eld.
FCC Appendix
Manufacturer’s declaration - electromagnetic emission
The WIRELESS IMAGE system is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the WIRELESS IMAGE system should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment -guidance
RF Emissions
CISPR 11
Group 1
The WIRELESS IMAGE system uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The WIRELESS IMAGE system is suitable for use in all establish-
ments other than domestic and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
A
Voltage fl uctuations
IEC 61000-3-3
Complies
Manufacturer’s declaration - electromagnetic immunity
This WIRELESS IMAGE system is intended for use in the electromagnetic environment specifi ed below. The
customer or the user of the WIRELESS IMAGE system should assure that it is used in such an environment.
Immunity test
IEC 60601 Test level Compliance level
Electromagnetic environment-guidance
Electrostatic
discharge(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or cera-
mic tile. If fl oors are covered with synthetic
material, the relative humidity should be
at least 30%
Electrical fast tran-
sient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/out-
put lines
2 kV for power
supply lines
1 kV for input/out-
put lines
Main power quality should be that of a ty-
pical commercial or hospital environment.
Surge
IEC 61000-4-5
1 kV diff erential mode
2 kV common mode
1 kV diff erential mode
2 kV common mode
Main power quality should be that of a typi-
cal commercial or hospital environment.
Power frequency
(50/60Hz) Magnetic
fi eld IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency magnetic fi elds should
be at levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.
Voltage dips, short
Interruptions and
voltage varia-
tions on power
supply input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycle
70%
U
T
(30% dip in
U
T
) for
25 cycle
<5%
U
T
(<95% dip in
U
T
)
for 5 sec.
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycle
70%
U
T
(30% dip in
U
T
) for
25 cycle
<5%
U
T
(<95% dip in
U
T
)
for 5 sec.
Main power quality should be that of a ty-
pical commercial or hospital environment.
If the user of WIRELESS IMAGE system re-
quires continued operation during power
mains interruptions, it is recommended
that WIRELESS IMAGE system be powered
from an uninterruptible power supply or
a battery.
Note:
U
T
is the A.C. mains voltage prior to application of the test level.