Maintenance of Quality of Service WUH4060
WUH4060
WUH4060 is designed to maintain an adequate Quality of Service during its use in a hospital
operating room environment.
Design Characteristics: Interference from other WUH4060 Devices
Using a unique mating procedure implemented during manufacture, a WUH4060 transmitter is
programmed to work only with authenticated WUH4060 receivers. WUH4060 has been tested to
work consistently within a 30m radius of the transmitter.
If more than one WUH4060 wireless systems are operating simultaneously in a single room, DFS
frequency characteristics avoid interference between systems.
It is important to physically separate all WUH4060 units, TX or RX, from each other by a minimum
distance of 1.525 meters or 5 feet.
As a result of these design considerations, it is not possible for one WUH4060 System to interfere
with another WUH4060 System.
Design Considerations: Interference from other Devices
The WUH4060 is validated to be compatible with other devices that are commonly found in an
operating environment, in accordance with IEC 60601-1-2. In that testing, the WUH4060 needed
to maintain the following Essential Performance to ensure that its performance was adequate
when the System was exposed to other electronic equipment. WUH4060 Essential Performance
requires that a viewable image shown on a secondary monitor via a WUH4060 pair be consistent
with the following criteria:
1. A viewable image is generated
2. There is no fl ickering of the image
3. There is no clipping of any edges
4. Brightness and detail of image remains consistent across the whole image
These criteria ensure that no resultant video degradation could be judged to be clinically signifi -
cant.