39
8/2021
Advanced TX and RX pairing if required.
NOTE:
All WUH4060 TX and RX units sold together are programmed at the factory, and should
connect automatically. For advanced unit pairing, see the steps below, or contact FSN Medical
Technologies for detailed information.
1.
Prepare Receiver (RX). Confi rm that the power on.
2.
Check the RX OLED display for the Lock/Unlock menu. If Locked - push and hold the joystick
left for 5 seconds to Unlock. Do the same to change Unlock to Lock.
3.
Enter RX OLED menu by pushing right on the joystick.
4.
Navigate the menu by pushing up or down on the joystick.
5. Select
the
UNPAIR
mode by pushing right on the joystick twice. Select
PROCEED TO UNPAIR
.
A message
UNPAIRING...PLEASE WAIT
is briefl y displayed.
6.
Next, select the
PAIR
mode by pushing right on the joystick once to start pairing. A message
PAIRING SEARCHING FOR TX
is displayed.
7.
Prepare Transmitter (TX). Confi rm that the power on.
8.
Check the TX OLED display for the Lock/Unlock menu. If Locked - push and hold the joystick
left for 5 seconds to Unlock. Do the same to change Unlock to Lock.
9.
Enter TX OLED menu by pushing right on the joystick.
10. Navigate the menu by pushing up or down on the joystick.
11. Select
the
PAIR
mode by pushing right on the joystick once. A message
PAIRING
is briefl y
displayed.
12. If successful, the TX OLED will display the message
PAIRING COMPLETED SUCCESSFULLY
.
13. If the OLED message shows
PAIRING FAILED,
repeat items 3 through 12.
Summary of Contents for FHD-1TX-1RX