Installation and Startup
Maxi-Cool
4
Rev 007, 04/12
© SP Scientific 2012
Setup
Physical Location
The area where you place the unit should be smooth, level, and capable of safely
supporting the weight of the machine. Space should be provided so that air can
properly circulate around the unit. If obstacles are blocking the airflow, the compressor
may overheat. This reduces performance, and can lead to early compressor failure.
The environment should be well ventilated to prevent excessive buildup of heat.
Normal room ventilation is usually adequate.
NEVER PLACE THE UNIT IN AND AREA WHERE EXCESSIVE HEAT, MOISTURE OR
CORROSIVE MATERIALS ARE PRESENT.
Ambient Air Requirements
This unit is designed for operation in a normal indoor environment. Systems should
not be mounted outside or otherwise exposed to the elements. The environment
should be free from air containing large amounts of moisture, salt or sulfur. If your
system utilizes an air-cooled condenser, the fan in the system draws ambient air
across the condenser (radiator) to cool the refrigerant.
Ambient Air Temperature Requirements
55
°
F to 69
°
F (12
°
C to 21
v
C)
Acceptable
70
°
F to 75
°
F (21
°
C to 24
°
C)
Ideal
75
°
F to 80
°
F (24
°
C to 27
°
C).
Acceptable, but some degradation of the cooling capacity of the
unit should be expected
80
°
F to 85
°
F (27
°
C to 29.5
°
C)
Expect reduced reliability. Warranty may be voided
Over 85
°
F (Over 29.5
°
C)
Not Acceptable.
Warranty will be Voided.
NEVER OPERATE THE MAXI-COOL IN AN AREA WITH AN AMBIENT TEMPERATURE ABOVE
32°C (90°F).
Water Requirements
If your unit is water cooled it will require 2-3 gpm of clean utility water at 30-60 psi., at
a temperature of 13°C to 22°C.
Summary of Contents for MAXI-COOL
Page 1: ...RECIRCULATING CHILLER OPERATOR S MANUAL MAXI COOL with EDC Control ...
Page 2: ......
Page 6: ...iv Rev 007 04 12 SP Scientific 2012 ...
Page 10: ...Introduction Maxi Cool 2 Rev 007 04 12 SP Scientific 2012 ...
Page 19: ...Maxi Cool Installation and Startup Rev 007 04 12 11 SP Scientific 2012 ...
Page 26: ...Operation Maxi Cool 18 Rev 007 04 12 SP Scientific 2012 ...
Page 36: ...User Menu Maxi Cool 28 Rev 007 04 12 SP Scientific 2012 ...
Page 65: ......