HOW TO USE
1. Press On/Off button then LCD displays
188.8
℉
M
(188.8
℃
M
)
for 2
seconds. Keep pressing then displays last measurement, and releasing
98.6℉(37.0℃)for 1 second with ℉or℃ flashing respectively.
2. If ambient temperature is higher than 111.2
℉
(44.0
℃
),LCD displays H
℉
(H
℃
)and lower than 89.6
℉
(32.0
℃
) then displays L0
℉
(L0
℃
).
3. Disinfect the probe with alcohol then place the probe under/in:
3.1 Oral use: Place the probe under tongue and close the mouth about
1minute till temperature is stable. Normal temperature is 98.6
℉
(36.7
℃
).
3.2 Axillary use: Wipe armpit with dry towel then place the probe in
armpit firmly about 1 minute till temperature is stable. Normal
temperature is 98.0
℉
(36.7
℃
).
3.3 Rectal use: Lubricate the probe with water-soluble jelly like K-Y
jelly. Do not use petroleum jelly like Vaseline for easier
insertion.
Insert the probe into rectum about 1 CM (lower than 1/2 inch)
till the temperature is stable. Normal temperature is about 99.7
℉
(37.6
℃
)
.
4. The peak temperature is first time and stable for 16 seconds, then alarm
Bi voice for 4 seconds is followed.
4.1 The power will be shut off automatically if no temperature rising
Notes:
1. Use the probe may result in 0.2
℉
(0.1
℃
)difference with real
temperature.
2. It is suggested that rectal use t---thermometer does not use for axillary
and oral use for the hygienic reason.
The Blue Tooth function model is BT-A41CN-BT
Operation Method:
★
Install the APK accordance with the communication protocol into
the blue tooth signal receiving device such as mobile phone.
★
Activate the blue tooth signal receiving device such as mobile
phone to match with the blue tooth of infrared thermometer