background image

 

- 11 -

4. Changes or modifications not expressly approved by the party responsible for compliance could 

void the user’s authority to operate the equipment. 

   This equipment has been tested and found to comply with the limits for a Class B digital device, 

pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection 
against harmful interference in a residential installation. This equipment generates, uses and can 
radiate radio frequency energy and, if not installed and used in accordance with the instructions, 
may cause harmful interference to radio communications. However, there is no guarantee that 
interference will not occur in a particular installation. 

If this equipment does cause harmful interference to radio or television reception, which can be 
determined by turning the equipment off and on, the user is encouraged to try to correct the 
interference by one or more of the following measures: 

-- Reorient or relocate the receiving antenna. 

-- Increase the separation between the equipment and receiver. 

-- Connect the equipment into an outlet on a circuit different from that to which the receiver is 

connected. 

-- Consult the dealer or an experienced radio/TV technician for help. 

5. Guidance and manufacturer’s delclaration 

 

Guidance and manufacture’s declaration – electromagnetic emission

The 

[EQUIPMENT or SYSTEM] i

s

 intended for use in the electromagnetic environment specified below. The customer of 

the user of the

 

 

[EQUIPMENT or SYSTEM] 

should assure that it is used in such an environment. 

Emission test 

Compliance 

Electromagnetic environment – 

guidance 

RF emissions 
CISPR 11 

Group 1 

The [EQUIPMENT or SYSTEM] use RF 
energy only for its internal function. 
Therefore, its RF emissions are very low 
and are not likely to cause any 
interference in nearby electronic 
equipment. 

RF emission 
CISPR 11 

Class B 

 

Harmonic emissions 
IEC 61000-3-2 

Not applicable 

Voltage fluctuations/ flicker 
emissions 
IEC 61000-3-3 

Not applicable 

 

Guidance and manufacture’s declaration – electromagnetic immunity 

The 

[EQUIPMENT or SYSTEM] 

is intended for use in the electromagnetic environment specified below. The customer 

or the user of 

[EQUIPMENT or SYSTEM] 

should assure that it is used in such an environment. 

Summary of Contents for FT-F31

Page 1: ...1 Instruction for Use lnfrared Thermometer No contact Type FT F31 BT FT F32 BT FT F41 BT FT F42 BT FT F31 FT F32 FT F41 FT F42...

Page 2: ...r Measuring Temperature 9 Product Specification 10 Common question concerning forehead temperature 2 Product Overview This infrared Thermometer measure the temperature It can get more quick and accura...

Page 3: ...hermometer and then all displayed segments appear briefly as like on Figure 5 2B the displayed screen appears last measured temperature data please see on Figure 5 3B after few seconds automatically c...

Page 4: ...ymbol appear on the display and the temperature unit or blink then it s ready to measure please see on Figure 5 4C Take IR thermometer to aim at the target about 5 to 10 CM distance then press the SCA...

Page 5: ...ll 12 memories recall for FT F31 FT F31 BT FT F32 FT F32 BT Press SCAN button to display M on the displayed screen in the power off condition and then memory recall displays Press again SCAN button ke...

Page 6: ...90 Isopropyl alcohol IPA to clean up before or after use Avoid multiple using by different peoples for cross infection Note 1 Any solution dilution and method other than mentioned above might cause r...

Page 7: ...ressure 86 106 kPa Storage Condition 25 0 55 0 13 0 131 0 Rh 95 Pressure 86 106 kPa Battery type and battery life AAA 2pcs Suitable for FT F31 FT F31 BT FT F42 FT F42 BT AA 2pcs Suitable for FT F41 FT...

Page 8: ...etooth 4 1 BLE Bluetooth Modulation Type GFSK Specially Voice Function for Model no FT F31 FT F31 BT FT F41 FT F41 BT 1 When the IR thermometer is ready you can hear Please measure voice or other lang...

Page 9: ...ermometer for the reading 3 Why is a forehead temperature is more accurate than ear temperature Ear thermometers are considered inaccurate because the positioning of the probe in the ear canal might a...

Page 10: ...anufacturer This device must only be serviced repaired and opened by individuals at authorized sales centers This device is not a life supporting me equipment Statement Clinical accuracy characteristi...

Page 11: ...etween the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for...

Page 12: ...nterruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5...

Page 13: ...arked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all sit...

Page 14: ...738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23 333 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estima...

Page 15: ...ng instructions IP22 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Management Zone Changping Town Dongguan...

Reviews: