Immunity test 

IEC 60601 test 

level 

Complian

ce level 

Electromagnetic environment - 

guidance 

Conducted RF 

IEC 61000-4-6 

3 V

rms

 

150 kHz to 80 MHz 

Not applicable

Portable and mobile RF communications 

equipment should be used no closer to any 

part of the 

[EQUIPMENT or SYSTEM]

, including 

cables, than the recommended separation 

distance calculated from the equation 

applicable to the frequency of the transmitter. 

Recommended separation distance 

 

P

d

167

.

1

=

          

 

P

d

167

.

1

=

         80 MHz to 800 MHz 

 

P

d

333

.

2

=

         800 MHz to 2.5 GHz 

Where 

P

 is the maximum output power rating 

of the transmitter in watts (W) according to the 

transmitter manufacturer and 

d

 is the 

recommended separation distance in metres 

(m). 

Field strengths from fixed RF transmitters, as 

determined by an electromagnetic site 

survey,

a

 should be less than the compliance 

level in each frequency range.

b

 

Interference may occur in the vicinity of 

equipment marked with the following symbol: 

 

 

 

 

Radiated RF 

IEC 61000-4-3 

 

3 V/m 

80 MHz to 2.5 GHz 

 

3 V/m 

NOTE 1        At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2        These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 

and reflection from structures, objects and people. 

a

         Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 

survey should be considered. If the measured field strength in the location in which the  

[EQUIPMENT or 

SYSTEM] 

is used exceeds the applicable RF compliance level above, the   

[EQUIPMENT or SYSTEM] 

should be 

observed to verify normal operation. If abnormal performance is observed, additional measures may be 

necessary, such as reorienting or relocating the   

[EQUIPMENT or SYSTEM]

b

        Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

Summary of Contents for FT-F31

Page 1: ...1 Instruction for Use lnfrared Thermometer No contact Type FT F31 BT FT F32 BT FT F41 BT FT F42 BT FT F31 FT F32 FT F41 FT F42...

Page 2: ...r Measuring Temperature 9 Product Specification 10 Common question concerning forehead temperature 2 Product Overview This infrared Thermometer measure the temperature It can get more quick and accura...

Page 3: ...hermometer and then all displayed segments appear briefly as like on Figure 5 2B the displayed screen appears last measured temperature data please see on Figure 5 3B after few seconds automatically c...

Page 4: ...ymbol appear on the display and the temperature unit or blink then it s ready to measure please see on Figure 5 4C Take IR thermometer to aim at the target about 5 to 10 CM distance then press the SCA...

Page 5: ...ll 12 memories recall for FT F31 FT F31 BT FT F32 FT F32 BT Press SCAN button to display M on the displayed screen in the power off condition and then memory recall displays Press again SCAN button ke...

Page 6: ...90 Isopropyl alcohol IPA to clean up before or after use Avoid multiple using by different peoples for cross infection Note 1 Any solution dilution and method other than mentioned above might cause r...

Page 7: ...ressure 86 106 kPa Storage Condition 25 0 55 0 13 0 131 0 Rh 95 Pressure 86 106 kPa Battery type and battery life AAA 2pcs Suitable for FT F31 FT F31 BT FT F42 FT F42 BT AA 2pcs Suitable for FT F41 FT...

Page 8: ...etooth 4 1 BLE Bluetooth Modulation Type GFSK Specially Voice Function for Model no FT F31 FT F31 BT FT F41 FT F41 BT 1 When the IR thermometer is ready you can hear Please measure voice or other lang...

Page 9: ...ermometer for the reading 3 Why is a forehead temperature is more accurate than ear temperature Ear thermometers are considered inaccurate because the positioning of the probe in the ear canal might a...

Page 10: ...anufacturer This device must only be serviced repaired and opened by individuals at authorized sales centers This device is not a life supporting me equipment Statement Clinical accuracy characteristi...

Page 11: ...etween the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for...

Page 12: ...nterruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5...

Page 13: ...arked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all sit...

Page 14: ...738 1 1 167 1 167 2 333 10 3 689 3 689 7 379 100 11 667 11 667 23 333 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estima...

Page 15: ...ng instructions IP22 Symbol for the IP classification Symbol for RF transmitters Manufacturer FUDAKANG INDUSTRIAL CO LTD Address No 8 Yinghe Road Yuanjiangyuan Management Zone Changping Town Dongguan...

Reviews: