SS1807-N002
Instruction of System
-
88/106
A) The image file format is Exif-JPEG or TIFF.
Accordingly, saved images can be displayed by using image viewer software.
B) Dedicated software for viewing the transferred image is not prepared at this time.
C) The EP-6000 does not have the function for viewing the images transferred to the FTP
server.
D) The rough estimation of the number of images to be recorded in the internal memory of
EP-6000 and the number of inspections is as follows.
Maximum
number of
recordable
images (for all
examinations)
Maximum
number of
recordable
images (for each
examination)
Standard number
of recordable
examinations
(100 images /
examination)
Minimum number
or recordable
examinations
(Maximum /
examination)
Compression
rate
JPEG
(1/20)
21,690
800
216
27
JPEG
(1/10)
16,270
750
165
21
JPEG
(1/5)
5,910
500
59
11
TIFF
840
130
8
6
* The number of images which can be recorded varies according to the photo subject.
* When a Full HD supporting monitor is connected and the screen is displayed in Full HD,
images are recorded in SXGA mode screen image.
◆
Note
◆
・
If there are non-transferred images, one of the reasons is considered as network abnormality
or image server abnormality.
It falis again if the images are still transferred with the abnormality.
Check that the network is connected correctly, the hub or router is working normally if it is
connected and check the image server is started normally in advance.
Be sure the network and server are normal and then transfer the image again. Check that
image is recorded correctly on the server.
・
If the images are not necessary though some images are not copied or the images that are
not transferred in backup, search and delete the images, or delete all the images by
initializing the built-in memory.
Must confirm there is no necessary image before deleting them.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...