7.2 Description
The EP-6000 has a network terminal (RJ45 terminal).
By connecting a twisted pair cable to the network terminal, the EP-6000 is connected to the FTP
server via the network.
7.2.1 Communication System
Protocol: TCP/IP
Application: File Transfer Protocol (FTP)
TheEP-6000 processor becomes an FTP client and sends data to the computer running an FTP
server service.
7.2.2 Connection
Connect to a network switch (switching HUB) or PC.
Depending on the connection condition, use the network switch (switching HUB) or PC
compliant with the safety standards listed below.
Safety standard
Cable type
Remarks
○
Outside the
patient
environment
Network switch
IEC60950
UL60950
Straight
×
PC
IEC60950
UL60950
Cross
×
Inside the
patient
environment
Network switch
EN60601-1
UL2601-1
Straight
No applicable
product
△
PC
EN60601-1
UL2601-1
Cross
Hard to get it
Configuration indicated with “○” is recommended.
[Note] Use a peer-to-peer connection.
[Note] Before connecting to the network, be sure to obtain authorization from the network
administrator.
[Note] After setting the network, turn off and then on the processor.
[Note] Do not communicate beyond the router.
Before using the network, installation of the FTP server application, settings on the server PC
(setting the network, creating a folder for image data, etc.) and settings on the EP-6000 are
required.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...