SS1807-N002
Failure Analysi
-
7/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
45
Information
Cannot be registered any
more.
Delete patient from
patient list.
When information of 45 patients
has been registered upon
pressing F1 on the observation
screen
Same as on the left
46
Information
Trigger mode is set to
"Freeze".
PinP mode is fixed to
"ON".
When the trigger mode was set
at “Freeze.”
Same as on the left
47
Information
Because the image to be
printed is not selected,
print preview screen
cannot be opened.
The start-up button was pressed
on the print preview screen
without images selected.
49
Information
Switching patient
information.
Displayed while the current
patient information was being
switched
Same as on the left
50
Information
Peripheral connecting
port duplicated.
Correct the settings.
When trying to set the peripheral
connecting port redundantly
51
Information
FICE settings saved.
At the time of FICE setting
completion
Press the Save button on the
FICE setting screen
52
Information
Registered as a new
patient.
OK?
When patients with appropriate
patient ID were not available at
the time of reading the magnetic
card on the observation screen
Same as on the left
In case of repetitive failures even
if the patients with appropriate
patient ID are available, replace
the magnetic card and the
magnetic card reader.
53
Information
Appropriate patient is not
found.
When patients with appropriate
patient ID were not available at
the time of reading the magnetic
card on the patient list screen
Same as on the left
In case of repetitive failures even
if the patients with appropriate
patient ID are available, replace
the magnetic card and the
magnetic card reader.
54
Critical error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
processor is in
abnormal state.
At the time of system start-up,
the image parameter files to be
read are not available, broken or
of obsolete version.
Same as on the left
Install a correct image parameter
file.
In case of repetitive failures after
such installation, replace the
main PCB.
55
Critical error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
processor is in
abnormal state.
When ALPS error occurred
Same as on the left
In case of repetitive failures,
replace the main PCB.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...