SS1807-N002
Failure Analysi
-
8/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
56
Critical error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
processor is in
abnormal state.
When VPFE error occurred
Same as on the left
In case of repetitive failures,
replace the main PCB.
57
Critical error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
processor is in
abnormal state.
Start-up was not completed
within the designated duration
(within 30 seconds after
application start-up)
30 seconds lapsed after
application start-up.
Wait for 10 seconds after power
was turned OFF, and then turn
ON the power.
In case of repetitive failures,
replace the main PCB.
60
Information
Patient information will be
cleared.
OK?
Pressed the contents-clearing
button on the patient information
input screen.
Same as on the left
61
Information
System data will be reset
to factory default.
OK?
“Factory Reset” was pressed on
the factory menu
Same as on the left
62
Information
Deleting information.
Please wait for a while.
Displayed at the time of deleting
all the patient and doctor
information
Same as on the left
63
Warning
Cannot be shot any more.
Cause: Number of
shots exceeds
upper limit.
End examination.
The number of shots exceeds the
upper limit, 999
Same as on the left
In case of repetitive failures even
when the number of shots has
not yet reached 999 (by counting
the number of shots as +1 when
shot with FICEOFF, +2 when
shot with FICEON, and +1 when
captured by switching the FICE
number in freeze state), replace
the main PCB.
64
Information
Remaining number of
shots is 100.
Because the remaining number
of shots has come to 100.
Same as on the left
65
Information
Image list display can be
used while the scope
button lights orange.
The key “Image List” was
pressed during examination.
Same as on the left
In case of repetitive failures even
if the scope button is lighted
orange, replace the main PCB.
66
Information
Search can be used while
the scope button lights
orange.
The key “Search” was pressed
during examination.
Same as on the left
In case of repetitive failures even
if the scope button is lighted
orange, replace the main PCB.
67
Information
Move the cursor to the
position where patient
information is registered.
When the magnetic card was
passed through, the patient list
was not focused on.
Same as on the left
In case of repetitive failures even
if the cursor is placed on the
position, replace the main PCB.
In case of repetitive failures even
if the magnetic card was not
passed through, replace the
magnetic card reader.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...