SS1807-N002
Caution in Safety
-
3/12
■ Handling products with suspicion of contamination at outside medical facilities
Maintenance should be implemented with particular attention to the following points:
・
If there is a local standard and/or regulation for pollution abatement, please comply with that
standard and/or regulation.
・
Wear protective clothing and gears during maintenance such as white gown, safety goggles,
mask and gloves.
・
Do not carelessly touch human body with gloved hand during maintenance. (Eyes need special
attention)
・
If there is special concern on health, change a field service engineer.
・
Wash hands thoroughly after maintenance.
・
Contaminated products such as protective equipments and tools after a maintenance should
be pack in plastic bags and sealed, and taken way (brought back).
・
If there is a suspicion of accident, contact a manager immediately.
■ Handling returned products from the market with suspicion of contamination.
Disinfect everything which has contact with products.
1) Preparation
・
Disposable work cloth such as white gown (to be worn)
・
Eye protector such as safety goggles (to be worn)
・
Disposable gloves (to be worn)
・
Disposable mask (to be worn)
・
Hand antisepsis such as WELPAS (for hand disinfection)
・
Ethanol (for disinfection)
◆
Precaution
◆
Fundamentally, use ethanol with 70% density.
・
Gauze etc. (for wiping the equipment)
・
Plastic sheet etc. (for covering under the workspace for wiping the equipment
)
2) Disinfecting procedure
・
Wear protectors such as white gown, safety goggles and gloves.
・
Lay plastic sheet under the work place where the equipment is to be disinfected.
・
Disinfect the target equipment with ethanol soaked gauze
.
・
Dispose used items such as gloves, mask, gauze and plastic sheet in separate plastic bags
(Precaution).
・
Disinfect hand with WELPAS etc. after washing.
◆
Precaution
◆
Dispose used items such as gloves, mask, gauze , plastic sheet as “infectious wastes”.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...