SS1807-N002
Failure Analysi
-
9/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
68
Warning
Settings are reset to
default because data
abnormality has been
detected. If necessary,
load the backup data.
If the error
appears each time
the system is
booted, hardware
may be faulty.
Contact the sales
agent or our
service
representatives.
Checksum error of FeRAM
Same as on the left
In case of repetitive failures,
replace the main PCB.
69
Critical error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
processor is in
abnormal state.
Acquisition of frame memory for
image saving failed.
At the time of trigger execution
In case of repetitive failures,
replace the main PCB.
70
Information
To enable reset values,
restart the system.
Brightness adjustment resetting
was completed.
Same as on the left
In case of displaying even if
factory resetting is not executed,
replace the main PCB.
72
Information
Freeze release timing on
PC monitor changed.
Restart the system.
Pressed Shft+Alt+J.
Same as on the left
73
Information
Select the type of
processing for doctor
page.
At the time of doctor page
selection
Under condition of doctor page
not being registered, enter on the
list.
74
Information
Select the type of
processing for FICE page.
At the time of FICE page
selection
Under condition of FICE page not
being registered, enter on the list.
75
Information
System data will be reset
to defaults (recommended
values).
OK?
“Reset to default” was pressed
on the factory default menu.
Same as on the left
76
Information
Obtaining worklist.
Please wait for a while.
Obtaining modality worklist.
Under condition of use of DICOM
examination list, press F2 key.
77
Information
Obtaining worklist.
Please wait for a while.
Obtaining modality worklist.
On examination list screen, press
the button ”Update List” and
enter.
78
Information
Specified doctor not
registered.
Settings at start-up
are used for doctor
and FICE pages.
79
Information
Specified procedure not
registered.
80
Information
Enter patient information.
Inspection started with patient
information not entered.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...