SS1807-N002
Failure Analysi
-
16/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
152
Information
Install Language display
can be used while the
scope button lights
orange.
153
Information
The INSTALLLANGAGE (install)
screen was started
The installation screen was
started
154
Information
The INSTALLLANGAGE (install)
screen was ended
The installation screen was
ended
155
Information
Installation will be
executed.
OK?
A software version was upgraded
on the software version upgrade
screen.
The button of software version
upgrade was pressed
156
Information
To enable setup values,
restart the system.
157
Information
Installation is in progress.
Wait for a while.
158
Error
Installation failed.
Turn off and on the power.
159
Information
Installation is completed.
Turn off and on the power.
160
Information
The external memory setting
screen was started
161
Information
The external memory setting
screen was ended
162
Information
Change of screen
resolution will take effect
at next startup.
163
Information
The screen resolution
setting has been set to
SXGA. This will take
effect after restart.
When the setting of output
resolution was "FullHD", the
shortcut key for forced switching
of resolution was pressed
164
Information
Remaining number of
shots is 50.
The remaining number of shots
became 50.
Same as on the left
165
Information
Please review the freeze
setting.
When setting the same function
to more than one switch was
attempted.
166
Information
To end examination,
cancel freeze mode to
display a live image.
167
Warning
Corruption of main memory
(overall) was detected.
At the time of memory check at
startup
Collect the log.
168
Warning
Corruption of main memory
(software version/DICOM
version) was detected.
At the time of memory check at
startup
Collect the log.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...