SS1807-N002
Failure Analysi
-
21/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
228
Information
Selected examination
data will be deleted.
OK?
A folder was selected on the
search screen and the D key was
pressed
Same as on the left
Dialog selection→
229
Information
Deleting selected
examination data.
Please wait for a while.
230
Warning
DICOM setting can be
used while the scope
button lights orange.
Processing to control start-up of
DICOM service setting screen
during examination
+
control
dialog display
Same as on the left
Nothing special
231
Warning
The server information
setting is available when
the scope button lights
orange.
Processing to control start-up of
server information setting screen
during examination
+
control
dialog display.
Same as on the left
232
Warning
Patient information
linkage cannot be used
on this screen.
The acquired data is
canceled.
Only an
observation
screen, a patient
information input
screen, and a
patient list screen
can be used.
When patient information linkage
is executed on non-permitted
screen.
Same as on the left
233
Warning
There is not enough
memory for the
examination. After exiting,
delete data from Internal
Memory.
If you continue,
the image cannot
be saved.
The capacity required for the set
compression ratio could not be
allocated in the capacity
allocation processing.
* Deletion could not be
completed due to data that has
been copied but not yet
transferred.
* Mounting has failed.
* Deletion has failed.
Inspection start/switching
patients
234
Information
Selected examination
data will be copied. OK?
235
Information
All examination data will
be copied. OK?
236
Information
Today's examination data
will be copied. OK?
237
Information
Copying selected
examination data.
Please wait for a while.
⇒
Do not remove
External Memory
while copying
data.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...