SS1807-N002
Failure Analysi
-
32/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
8010
Information
Please check if the scope,
which was checked its
cleaning state last, is the
one currently attached
before starting the exam.
Inspection was started after
several times of checking the
state of cleanliness
8011
Information
Querying the server for
patient information.
Please wait for a while.
8012
Warning
Querying for patient
information will be
cancelled because the ID
has not been entered.
Please enter the patient's
ID before querying.
Patient's ID has
not been entered
yet.
Patient's ID has not been entered
yet
The patient information linkage
function utilization setting in the
server information setting (for
service) was "Use search of
patient ID only", and a patient ID
was not entered when requesting
patient information
Request patient information after
entering a patient ID.
8013
Warning
Querying for patient
information will be
cancelled because the
patient's ID cannot be
read.
Please check the
magnetic card/barcode
settings.
The patient's ID
cannot be read
from the magnetic
card/barcode.
Patient's ID has not been
registered yet
The patient information linkage
function utilization setting in the
server information setting (for
service) was "Use" or "Use
search of patient ID only", and an
patient ID was not registered
when requesting patient
information
Using barcode data with patient
ID registered, RFID data, make a
request of obtaining patient
information.
8014
Warning
A timeout error occurred.
Please check the timeout
value in the server
information setting.
When connection confirmation
failed.
Wrong setting of timeout, etc.
When connection confirmation is
completed after the
"Communication Test" button is
pressed on the server
information setting screen (for
service). (failure route)
Re-check timeout setting, etc.
8015
Information
Checking the connection
to the server.
Please wait for a while.
Screen checking the connection
Press the "Communication Test"
button on the server information
setting screen (for service).
Nothing special
8016
Information
Connection to the server
succeeded.
In case of success of connection
confirmation.
When connection confirmation is
completed after the
"Communication Test" button is
pressed on the server
information setting screen (for
service). (success route)
Nothing special
8017
Warning
Connection to the server
failed.
Check the setting.
When connection confirmation
failed.
Wrong setting of server, etc.
When connection confirmation is
completed after the
"Communication Test" button is
pressed on the server
information setting screen (for
service). (failure route)
Check server setting again.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...