SS1807-N002
Failure Analysi
-
33/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
8018
Warning
Patient information could
not be obtained from the
server.
8019
Warning
A failure occurred while
integrating
C@RNACORE.
Please check if the
network is in normal state.
8020
Warning
-
-
-
-
-
8021
Error
Communication Error.
Check Peripheral Status.
Response of inspection start
time-out
8022
Error
Communication Error.
Check Peripheral Status.
Response of inspection start
NAK
8023
Error
Communication Error.
Check Peripheral Status.
Response of capture time-out
8024
Error
Communication Error.
Check Peripheral Status.
Response of capture NAK
8025
Error
Communication Error.
Check Peripheral Status.
Response of scope information
time-out
8026
Error
Communication Error.
Check Peripheral Status.
Response of scope information
NAK
9001
Warning
Storage authentication
failed.
Storage functions cannot
be used.
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
9002
Information
Storage authentication
succeeded.
Storage functions can be
used.
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
9003
Warning
MWM/MPPS
authentication failed.
MWM/MPPS functions
cannot be used.
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
9004
Information
MWM/MPPS
authentication succeeded.
MWM/MPPS functions
can be used.
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...