SS1807-N002
Failure Analysi
-
34/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
9005
Warning
Since authentication of
Storage and MWM/MPPS
failed, DICOM cannot be
used.
DICOM utilization setting
changed to "Not used".
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
9006
Warning
Authentication of Storage
and MWM/MPPS
succeeded.
To use DICOM functions,
change DICOM function
utilization setting.
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
9007
Information
Checking the connection
to the server.
Please wait for a while.
Screen for checking the
connection
On DICOM setting screen (for
service), press the
button ”Verify.”
Nothing special
9008
Information
Connection to the server
succeeded.
In case of success of connection
confirmation
When connection confirmation
was finished after the
button ”Verify” was pressed on
DICOM setting screen (for
service) (success route)
Nothing special
9009
Warning
Connection to the server
failed.
Check the setting.
In case of failure of connection
confirmation.
Wrong setting of server, etc.
When connection confirmation
was finished after the
button ”Verify” was pressed on
DICOM setting screen (for
service) (failure route)
Check server setting again.
9010
Warning
Authentication status
changed as follows.
Storage license
authentication - Failed
MWM/MPPS license
authentication -
Succeeded
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
9011
Warning
Authentication status
changed as follows.
Storage license
authentication -
Succeeded
MWM/MPPS license
authentication - Failed
When mismatch occurred
between the results checked in
advance on the setting screen
and the results checked at the
time of setting completion
On DICOM setting screen (for
service), press the button ”Save
and Exit” and enter.
Check again if there is any
problem as the result of such
mismatch.
9012
Information
Creating DICOMDIR file.
Please wait for a while.
DICOMDIR index file creation
screen
When a DICOM file using a
trigger is created on the external
memory?
Nothing special
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...