SS1807-N002
Failure Analysi
-
40/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
20025
Information
Obtaining scope color
information.
Please wait for a while.
The scope color adjustment
screen was launched on a
400-series scope.
Same as on the left. The
message only appears for the
first time after the start of
inspection.
In case when the scope color
auto adjustment screen
frequently fails to open even after
waiting for a while, replace the
scope, the main PCB, and the
patient PCB.
20026
Information
Writing color settings of
530 Complementary
Color scope.
Please wait for a while.
Generation of 530 TV scope data
was executed.
Same as on the left
In case when the generation of
the 530TV scope data frequently
fails to complete even after
waiting for a while, replace the
scope, the main PCB, and the
patient PCB.
20027
Error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
endoscope is in
abnormal state.
A sensor initial communication
error occurred in the S600 scope.
At notification from the S600
scope
In case of repetitive failures,
replace the S600 scope.
20028
Error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
endoscope is in
abnormal state.
A sensor communication error
occurred in the S600 scope.
At notification from the S600
scope
In case of repetitive failures,
replace the S600 scope.
20029
Error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
endoscope is in
abnormal state.
An EEPROM device error
occurred in the S600 scope.
At notification from the S600
scope
In case of repetitive failures,
replace the S600 scope.
20030
Error
Turn off processor, unplug
and plug back scope, and
then turn it back on. If the
problem persists, contact
our service
representatives.
Cause: A power
supply error
(over-current)
occurred.
A power supply error
(over-current) occurred in the
S600 scope.
At notification from the S600
scope
In case of repetitive failures,
replace the S600 scope.
20032
Warning
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
endoscope
currently
connected is
disabled by the
setting.
Elle scope was connected while
the use of Elle was disabled.
20033
Information
Scope log file will be
exported in External
Memory.
OK?
Exporting Scope log file was
selected at the backup of Service
menu.
Same as on the left
Dialog selection
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...