SS1807-N002
Failure Analysi
-
41/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
20034
Information
Exporting Scope log file in
External Memory.
Please wait for a while.
⇒
If the 700 siries
scope is not
connected, the
light source unit
and the 700 sireis
scope log file
cannot be
exported.
”Yes” was selected at the above
dialog.
Same as on the left
Wait till processing is finished. If
the message does not disappear
after a while, replace External
Memory, USB PCB, and the main
PCB.
20035
Information
Exporting Scope log file is
completed.
Exporting Scope log file was
completed.
Same as on the left
Press any key to close the
dialog.
20036
Warning
Exporting Scope log file
has failed.
Check External Memory.
Exporting Scope log file has
failed.
・
Abnormality in getting the
current time and date.
・
File copying to CF failed.
Same as on the left
After the contents of the inserted
CF have been backed up in
Windows PC, etc., insert CF
again and export the file after CF
has been physical-formatted
instead of being quick-formatted.
In case of repetitive errors after
that, replace CF, CF PCB and the
main PCB.
20037
Warning
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: Light
source unit or
endoscope is
connected
improperly.
Failure of the patient PCB or
poor connection with faulty
connection with patient PCB at
the time of communication
processing with the patient PCB.
or
Failure of the main PCB, flexible
flat cable, or patient power supply
At the time of communication
with the patient PCB
Consult your local FUJIFILM
dealer, or the service center from
which you purchased the
product. In case of repetitive
failures, replace the main PCB
and the patient PCB.
20038
Warning
Connect or disconnect the
endoscope when the
EXAM. button lights
orange or when the power
is OFF.
The scope was removed during
examination.
Same as on the left
In case of repetitive failures in
spite of the scope not being
removed, replace the scope
connector and the scope.
20039
Error
Clean the equipment in
the proper way as
specified.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: Light
source unit or
endoscope is
connected
improperly.
Detection of foreign matter
(safety circuit drive)
Error message: 0x0001
20040
Error
Unplug and plug back
endoscope, reset the
processor. If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The
endoscope is in
abnormal state.
(SCPU) Abnormality with LD
driver
Error message: 0x0010
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...