SS1807-N002
Failure Analysi
-
50/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
30137
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: LD low temperature error
was detected in light source.
Brightness control was activated
and re-lighting was inhibited.
At detection of temperature error
in LL-7000 light source
Wait until LD temperature is
raised.
Replace the light source if not
improved.
30138
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: LD high temperature
error was detected in light
source.
Brightness control was activated
and re-lighting was inhibited.
At detection of temperature error
in LL-7000 light source
Wait until LD temperature is
lowered.
Replace the light source if not
improved.
30139
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
(CODE =
0x11203)
Cause: Heat sink low
temperature error was detected
in light source.
At detection of temperature error
in LL-7000 light source
Wait until LD temperature is
raised.
Replace the light source if not
improved.
30140
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
(CODE =
0x11204)
Cause: Heat sink high
temperature error was detected
in light source.
At detection of temperature error
in LL-7000 light source
Wait until LD temperature is
lowered.
Replace the light source if not
improved.
30141
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: Cooling fan stop error
was detected in light source.
This is only a warning and
operation is not limited.
At detection of hardware error in
LL-7000 light source
No emergency response is
required.
(It will eventually become LD
temperature error.)
30143
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Code error of LD PCB (excluding
standard LD drive PCB) was
detected at light source startup.
This is only a warning and
operation is not limited. However,
normal operation is not ensured.
At detection of hardware error in
LL-7000 light source
Replace the light source.
30144
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: EEPROM readout error
was detected in light source.
This is only a warning and
operation is not limited. However,
normal operation is not ensured.
At detection of hardware error in
LL-7000 light source
Replace the light source.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...