SS1807-N002
Failure Analysi
-
51/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
30145
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: EEPROM writing error
was detected in light source.
This is only a warning and
operation is not limited. However,
normal operation is not ensured.
At detection of hardware error in
LL-7000 light source
Replace the light source.
30146
Information
Checking the status of the
built-in light source
element.
Cause: The light
source LL-7000 is
in abnormal state.
A failure was detected in
semiconductor laser device with
built-in light source.
The faulty channel is under
investigation.
At detection of LD unit failure in
LL-7000 light source
Replace the light source.
30147
Warning
Operating in emergency
lamp mode. Take an
appropriate measure
referring to the operation
manual and contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: 405nmLD failure was
detected in light source.
At detection of LD unit failure in
LL-7000 light source
Replace the light source.
30148
Warning
Operating in emergency
lamp mode. Take an
appropriate measure
referring to the operation
manual and contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: 445nmLD failure was
detected in light source.
At detection of LD unit failure in
LL-7000 light source
Replace the light source.
30150
Error
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: Overall LD failure was
detected in light source
At detection of LD unit failure in
LL-7000 light source
Replace the light source.
30151
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: Error of machine
difference data was detected in
light source.
Operation is continued with
outgoing inspection data.
At detection of data check error
in LL-7000 light source
Execute a reset.
If not improved, data is
determined to be corrupted.
30152
Warning
Unplug and plug back
endoscope, reset the
processor and the light
source.If the problem
persists, contact your
local FUJIFILM dealer.
Cause: The light
source LL-7000 is
in abnormal state.
Cause: Outgoing inspection data
error was detected in light
source.
Operation is continued with
default data.
At detection of data check error
in LL-7000 light source
Execute a reset.
If not improved, data is
determined to be corrupted.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...