SS1807-N002
Failure Analysi
-
64/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
43014
Error
Video recorder is in
abnormal state.
Check the device.
Error occurred in the video
recorder. Or a failure occurred in
the video recorder.
When response from the video
recorder is received in error
code.
In case of repetitive failures,
replace the video recorder.
44001
Error
Communication error with
DICOM GW equipment.
Check the connection.
Serial communication with DG
equipment is in abnormal state.
Same as on the left
In case of repetitive failures,
replace DG equipment and the
main PCB.
44002
Error
A communication error is
detected with SU
equipment.
Check the connection.
Error occurred in the serial
communication with ultrasonic
endoscope. Or a failure occurred
in the cable or the ultrasonic
endoscope.
Switching patients
Check the connection cable and
replace the main PCB in case of
repetitive failures.
44003
Information
Communication with the
ultrasonic processor was timed
out.
When ultrasonic processor is not
connected by serial
communication cable
*No message is displayed.
Logging only.
Check the cable connection with
ultrasonic processor.
Check the power of ultrasonic
processor.
Replace the cable.
Replace the light source if not
improved.
50005
Information
Please enter a password.
(Press [ESC] key to
cancel. )
Login authentication password
dialog
When the function needing login
authentication is selected
50006
Warning
New password and
Confirm password
mismatch.
Password was not matched.
When login password is input
50007
Warning
This Password is not
available.
Password was invalid.
When login password is input
50008
Warning
Incorrect Password.
Password was incorrect.
When login password is input
50009
Information
Changing password
succeeded.
Password was successfully
changed.
When login password is input
50010
Information
Login was successful: User
At successful login
50011
Information
Login was successful:
Administrator
At successful login
50012
Information
Login was successful: Service
At successful login
50013
Information
Login was successful: Service
center
At successful login
50014
Information
Login was successful: Factory
At successful login
50015
Information
Login was successful: Factory
base
At successful login
50016
Information
Log-off was executed.
At the time of logoff
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...