SS1807-N002
Checkup, Replacement and Adjustment
-
119/137
No.
Inspection Item
Acceptance Criterion
Inspection Method
NOTE
26
Brightness
adjustment
・
The brightness can be adjusted correctly.
・
The light limit functions properly.
・
Check that the indicator display changes from the
minimum to the maximum according to object
type and the distance to object. Also make sure
the brightness of the displayed image can be
controlled in a certain range and the response
time is within 1 second.
・
Assign a light limit to the MULTI button and press
the MULTI button, then check the indicator lights
up and “L-Limit” is displayed on the screen.
・
Connect a scope to let the lamp light up, and
check the scale marks of the indicator light up
no more than 7.
・
Press the MULTI button again, check that the
display of “L-Limit” disappears, and the brightness
and the indicator status revert to the first.
・
Revert the function assignment of MULTI button.
Jig / Equipment
・
700 system scope
27
Maximum output
light quantity
・
The output light quantity from the light source for
the normal observation mode is within the
standard range.
・
Connect a 700 system scope to the scope socket
and press the EXAM button to set the status to
ACTIVE.
・
Turn OFF the pump.
・
Press the EXAM button to set the status to
STANDBY, remove the scope and connect the
light measurement jig.
・
Measure the maximum light quantity for the
normal observation mode and check if it is within
the standard range written to [Table8].
Clean the window
glass of the jig with
alcohol before
measuring the light
quantity.
Jig / Equipment
・
620Y200004*
・
Y10N100217
28
PUMP button
・
By pressing the PUMP button, the air supply lamp
is switched in order as: “H” → “M” → “L” → OFF
・
Press the PUMP button and check if the air
supply lamp is switched.
29
Pump air supply
pressure/
Pump air supply
volume
・
The air supply pressure and air supply volume are
within the standard range written in the [Table9].
・
Measure the air supply pressure and air supply
volume with the measurement jig.
* Perform the measurement after 10 minutes since
the power has been turned ON.
Jig / Equipment
・
JA5003
・
Y11N100039
30
Air guide connection
・
No air leakage around the connection between
the rubber seal of scope socket and the LG
connector air guide.
・
Set the air supply pressure/volume measurement
jig on the scope socket, press the PUMP button to
switch the pump to “H”, and turn the jig in both
clockwise and anticlockwise direction to check
that the pressure decreases within 5kPa.
・
Turn OFF the pump and check the pressure does
not go down to zero within 5 seconds.
Jig / Equipment
・
JA5003
・
Y11N100039
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...