41S_Rev00
33
Basic Operation
“Good” message
Arc power and splicing position calibration are successfully completed. Press ESC key
to exit.
“Not Adequate” message
Arc power and splicing position calibration are completed but further calibration is
strongly recommended, as the change from the previous arc calibration is too large.
Press ENT key to perform arc calibration, or ESC key to exit even though arc calibration
is not completed.
Result: Good
In some cases, multiple iterations of arc calibration are needed until the
calibration process is successfully completed and the "Test Finish" message
is displayed. Arc calibration can be considered almost completed if multiple
iterations are completed without receiving the message.
Number threshold can be set so that "Test Finish" message is displayed after
specific number of arc calibrations is performed. See section [Maintenance
Settings] for detail.
Result: Not Adequate