62S_Rev3
86
Maintenance
“Good” message
Arc power and splicing position calibration are successfully completed. Press ESC key
to exit.
“Not Adequate” message
Arc power and splicing position calibration are completed but further calibration is
strongly recommended, as the change from the previous arc calibration is too large.
Press ENT key to perform arc calibration, or ESC key to exit even though arc calibration
is not completed.
Result: Good
In some cases, multiple iterations of arc calibration are needed until the calibration
process is successfully completed and the "Test Finish" message is displayed. Arc
calibration can be considered almost completed if multiple iterations are completed
without receiving the message.
Number threshold can be set so that "Test Finish" message is displayed after specific
number of arc calibrations is performed. See section [Machine Settings] for detail.
Result: Not Adequate