70S+A_Rev1
107
Error Message List
Error
Message
Reason
Solution
Cannot detect
fiber in
AUTO mode
・
Dust or dirt is on the fiber
surface.
・
Completely prepare the fiber again (strip, clean
and cleave).Select the suitable splice mode.
・
The left and right fiber
types are different.
・
Splicing non-standard
fibers.
Strong
Arc Power
・
Unable to calibrate due
to strong arc.
・
Replace the electrodes by using the [Replace
Electrodes] function in. If this does not
eliminate the problem, consult the authorized
distributor.
Weak
Arc Power
・
Unable to calibrate due
to weak arc.
Too Left /Right
Arc
・
Unable to calibrate due
to poor arc field position
(too far left or Right).
Fiber
Separation
・
The fiber stuff amount is
insufficient.
・
Execute the [Motor Calibration] function. If
using other splice modes, check the [Overlap]
setting in the splice mode.
・
The pre-fuse power or
pre-fuse time is set too
high.
・
Check the [Prefuse Power] and [Prefuse Time]
settings in the splice mode.
L/R-Bad Fiber
Position
・
The fiber is not set
correctly at the bottom of
the V-groove.
・
Press RESET key, and re-position the fiber
again to seat it correctly at the bottom of the
V-groove.
No Arc
Discharge
・
Arc Discharge did not
occur.
・
Assure the electrodes are in proper position.
・
Replace the electrodes.
・
Consult the authorized distributor.