Basic Operation
Symptom
Cause
Remedy
Separation
Fiber stuffing too
small
Perform [Motor Calibration]
Prefuse power too
high or prefuse time
too long.
Decrease [Prefuse Power] and/or [Prefuse
Time].
Fat
Fiber stuffing too
much
Decrease [Overlap] and perform [Motor
Calibration].
Thin
Arc power not
adequate
Perform [Arc Calibration].
Some arc parameters
not adequate
Adjust [Prefuse Power], [Prefuse Time] or
[Overlap].
Line
Some arc parameters
not adequate
Adjust [Prefuse Power], [Prefuse Time] or
[Overlap].
A vertical line sometimes appears at the splice point when MM fibers or dissimilar
fibers (including different diameters) are spliced. This does not affect splice quality,
such as splice loss or tensile strength.
A loss becomes high when the following errors occur.
Electric discharge may become unstable when usage
environment changes significantly.
In such a case, please use it after performing
“Stabilize Electrodes”.
When ENT key has been pushed, "Stabilize
Electrodes" is performed.
Please perform an "Arc Calibration" after the end
of "Stabilize Electrodes" according to the
directions on a screen.