70S+A_Rev1
48
Basic Operation
Wind-protector motion and sheath clamp motion
This unit has the Wind-Protector automatic opening-and-closing mechanism and switching
action differs by the chosen operation mode.
Operation
Mode
Wind-protector motion
When error
occurred such as
Fat,Thin,Loss etc.
When
Error not
occurred
When error
occurred except for
having enumerated
to the following
When the
power
supply
starts
When RESET
key is pushed
end of
splicing
Pause 1
end of
splicing
Auto-L
Auto-R
Auto-LR
closed
open
Fast
Normal
closed
closed
open
open
Norm-L
Norm-R
Norm-LR
Manual
closed
Customized
closed
open/closed
* When the power supply is off, "Wind-Protector" state is closed.
So please be sure to push a power button and to turn off a power supply.
Since "Wind-Protector" will open depending on the operation mode, we recommend you to set
the operation mode as Normal mode if you want to use additional electric discharge etc. after
end of splice.
Time after setting a fiber until "Wind-Protector" closes is short by "Auto" and "fast" mode, and is
set up a little long by "Normal" mode.
Time after setting a fiber until "Wind-Protector" closes can be changed in "customized" mode