70S+A_Rev1
70
Splice Menu
A continuation of
edit parameter list
Parameter
Description
Ignore Splicing Error
Loss
Set to “Enable” or “Disable”.
Enable
: An operator can accept and bypass an error shown on the left
by pressing the SET key.
Disable
: An operator can’t bypass an error shown on the left.
Bubble
Fat
Thin
Dust Burned
Cleave Angle
Set to “Enable” or “Disable”.
Enable
: An operator can accept and continue with the error as shown on
the left by pressing the SET key.
Disable
: An operator can’t override the error as shown on the left. The
unit repeats the alignment when the operator presses the SET
key.
Cleave Shape
Fiber Angle
Others
Max Num. of
Rearcs
The re-arcing process sometimes improves the splice loss. However,
excessive re-arcing may decrease the splice strength. With this function, it
is possible to limit the number of re-arcs or to disable re-arc discharge.
Cleaver Counter
Mode
When splicing, the splice increments the “Cleave Count”.