Maintenance
Operation procedure
1. Select [Arc Calibration] in [Maintenance
Menu] to display Arc Calibration screen.
2. Set prepared fibers onto the splicer.
3. The ARC Calibration will begin after pressing ENT key.
“Good” message
Arc power and splicing position calibration are successfully completed. Press ESC key
to exit.
Use standard SMF ITU-T G652 fiber for Arc Calibration.
Use well prepared fibers for arc calibration. Dust on the fiber surface affects arc
calibration.
Cleave angle threshold does not link to the parameter "Cleave Limit" in splicing
modes. Cleave angle threshold is independently set for arc calibration. See section
[Machine Settings] to change cleave angle threshold.
Result: Good