6.3 Remedying Typical Troubles
fi-6140 / fi-6240 Image Scanner Operator's Guide
6.3 Remedying Typical Troubles
Symptom: The scanned images are elongated when scanned from the ADF.
Are the rollers dirty?
YES
Clean the dirty rollers.
(For how to clean the rollers, see Section "
NO
Is any of consumables (Pick rollers and Brake
rollers) worn out?
YES
Check the consumables counters. Replace the
consumables, if needed.
(For how to replace the consumables, see
Chapter "
.")
NO
Have you adjusted the scan scale?
NO
Display the properties of the scanner that you
are using on "Scanners and Cameras" in the
control panel, then adjust the scan scale.
YES