17
CAUTION
Precautions for Use of SD-5C(20), SD-5C(25)
·
I
f measurements are taken over a long period of time, change the measuring site
several times a day to prevent rash or low-temperature burn.
·
When fixating the probe by bandage tape, do not tighten it too hard as this may
cause edema or congestion.
·
Since the skin of a neonate is extremely sensitive, be careful when removing the
adhesive tape.
·
Check that the LED (light source) and sensor are properly aligned . The sensor
must be on the opposite side of the LED.
·
When removing the probe from the patient or removing the bandage tape from
the probe, hold the probe pad and not the cable.
·
Cover the probe with a black cover if it is exposed to direct sunlight. Some
measurement error may occur if the probe is exposed to direct sunlight during
measurements.
·
Do not apply excessive force on the LED and sensor, otherwise damage to the
probe may result.
·
Inaccurate measurements can be caused by the placement of SpO
2
probe on an
extremity with a blood pressure cuff, arterial catheter, or intravascular line.