Operational Troubleshooting Flow Chart
Power On Pre Purge
No power:
Not plugged in
Surge Protector Faulty
Alarm light on:
Reset Burner Control
Pre Purge Ignition Transformer Engages
No Ignition:
Faulty spark plug Clean or replace
Faulty Boot or Connector Repair
Damaged Ignition cord
Replace
Faulty Transformer Replace
Ignition Transformer Engages
low fire Solenoid opens
No Flame:
Faulty/Blocked Solenoid
Replace/ Clean
Gas Bottle Empty Check at test point for flow. Change/fill bottle
Low gas pressure Check supply pressure at test point. Must be 140KPA.
If not then may be faulty Regulator.
Blocked Gas Filter
Remove and Clean
Blocked Inspirator jet Remove and Clean
low fire Solenoid opens
UV Flame detection
Flame Ignites but goes out:
Ignites for 5 seconds and goes out Check Uv Cell alignment to pilot flame
Check Uv Cell Globe for dust build up
Replace Uv Cell, may be faulty