The amplifier functionality can be tested on its own 
by confirming the following:

1.  An open or the sensor disconnected from the 

sensor side will will cause the amplifier to saturate 
to full output or higher. Amplifier should be 
powered with 12.5 VDC to 26 VDC through Pin 1 
(+Vin) and Pin 2 (GND)

2.  Connection of Pin 2 (+ Signal) and Pin 3 (– Signal) 

on the sensor side will cause the amplifier to go 
close to zero output condition.

 

Amplifiers with an offset induced using the amplifi-
er dipswitches would see the zero output condition 
being the offset value. For example 5 V for IAA100 
and 12 mA for the IAA200

All connections to amplifier should be made with the power first off. 
Connection to power with power on can damage amplifier.