The IHH500 is set to recognize TEDS automatically upon startup if this 
function is available.

This function must be enabled under the TEDS DATA menu.

Sensor cable shield connections should be grounded on one end, either 
the sensor side or the instrument side, to avoid potential ground loops.

Shield should be connected to cable clamp of binder cable assembly when 
connecting on instrument side.

For more information on TEDS see section 6.8 of the IHH500 manual.

More information on setting up and editing a sensor profile can be found 
in section 6.1 of the IHH500 manual.

NOTES

Sensor Solution Source

Load · Torque · Pressure · Multi-Axis · Calibration · Instruments · Software

www.futek.com

IHH500 Profile Setup Guide

G

M

F E

D

C

L

B

A

K

J

H

Output

Sensor

USB

Power

SENSOR CONNECTIONS

PIN

SYMBOL

DESCRIPTION

A

+E

+Excitation

B

+S

+Signal

C

–E

–Excitation, TEDS Return

D

–S

–Signal

E

TEDS_IO

TEDS Data

F

24_OUT

24V Output

PIN

SYMBOL

DESCRIPTION

G

GND_OUT Ground

H

5_OUT

5V Output

J

–V

–V and –mA Amplified Input  

Connections

K

+V

+V and +mA Amplified Input  

Connections 

L

PLEAD

Leading pulse from sensor

M

PLAG

Lagging pulse from sensor