The IHH500 is set to recognize TEDS automatically upon startup if this
function is available.
This function must be enabled under the TEDS DATA menu.
Sensor cable shield connections should be grounded on one end, either
the sensor side or the instrument side, to avoid potential ground loops.
Shield should be connected to cable clamp of binder cable assembly when
connecting on instrument side.
For more information on TEDS see section 6.8 of the IHH500 manual.
More information on setting up and editing a sensor profile can be found
in section 6.1 of the IHH500 manual.
NOTES
Sensor Solution Source
Load · Torque · Pressure · Multi-Axis · Calibration · Instruments · Software
www.futek.com
IHH500 Profile Setup Guide
G
M
F E
D
C
L
B
A
K
J
H
Output
Sensor
USB
Power
SENSOR CONNECTIONS
PIN
SYMBOL
DESCRIPTION
A
+E
+Excitation
B
+S
+Signal
C
–E
–Excitation, TEDS Return
D
–S
–Signal
E
TEDS_IO
TEDS Data
F
24_OUT
24V Output
PIN
SYMBOL
DESCRIPTION
G
GND_OUT Ground
H
5_OUT
5V Output
J
–V
–V and –mA Amplified Input
Connections
K
+V
+V and +mA Amplified Input
Connections
L
PLEAD
Leading pulse from sensor
M
PLAG
Lagging pulse from sensor