IHH500 Profile Setup Guide
3
b.
Using the red left and red right arrow keys select the digit to
be changed. Using the red up and red down arrow keys set
the value for the selected digit that corresponds to the sensors
sensitivity (+) and use the red ENTER button to save the value.
•
The decimal place can be set by first changing the decimal
to a number and then changing another selected digit into
a decimal.
16. Setup SENSITIVITY (-)
a.
Using the red down arrow key followed by the red Enter button
select SENSITIVITY (-).
•
SENSITIVITY (-) can only be selected if BI- DIRECTIONAL
was selected earlier.
•
The sensitivity (-) will be the negative mV/V output for
bridge type sensors and the maximum negative voltage
output for amplified sensors.
•
The output of the sensor is usually listed on the sensor
calibration certificate.
•
For current output sensors, +/- Sensitivity is not available.
Calibration must be used for current output sensors.
b.
Using the red left and red right arrow keys select the digit to
be changed. Using the red up and red down arrow keys set
the value for the selected digit that corresponds to the sensors
sensitivity (-) and use the red ENTER button to save the value.
•
The decimal place can be set by first changing the decimal
to a number and then changing another selected digit into
a decimal.
17. Using the red down arrow key followed by the red Enter button select
CALIBRATION if a live calibration, where a load is to be placed on
the sensor, is desired. Otherwise, CALIBRATION can be skipped by
pressing the red down arrow key.
a.
Press the red Enter button to select Zero load (+) and proceed to
the input screen.
b.
With no load on the sensor, press the red Enter button to capture
the Zero load (+) reading.
c.
Press the red down arrow key followed by the red Enter button to
select FullScale (+) and proceed to the input screen.
d.
With the full load on the sensor, which results in the full positive
output, press the red Enter button to capture the FullScale (+)
reading.
e.
Press the red down arrow key followed by the red Enter button to
select Zero load (-) and proceed to the input screen.
•
The direction must be set to Bi-Directional for a – input in
CALIBRATION.
f.
With no load on the sensor, press the red Enter button to capture
the Zero load (-) reading.
g.
Press the red down arrow key followed by the red Enter button to
select FullScale (–) and proceed to the input screen.
•
The direction must be set to Bi-Directional for a – input in
CALIBRATION.
h.
With the full load on the sensor, which results in the full negative
output, press the red Enter button to capture the FullScale (-)
reading.
18. Using the red down arrow key followed by the red Enter button select
PULSE / ROTATE.
a.
The PULSE / ROTATE menu is selectable only for profiles that
specify PULSE as the sensor type.
b.
Typical FUTEK settings are 360 and 60.
19. Using the red left and red right arrow keys select the digit to be
changed. Using the red up and red down arrow keys set the value for
the selected digit that corresponds to the encoder Pulse counts. Use
the red ENTER button to save the value.
a.
The default value is set to 360.
b.
See section 6.1.3.8 of the IHH500 manual for more information
on PULSE / ROTATE.
20. Using the red down arrow key followed by the red Enter button select
SERIAL NUMBER.
21. Using the red left and red right arrow keys select the digit to be
changed. Using the red up and red down arrow keys set the value for
the selected digit that corresponds to the sensor’s serial number. Use
the red ENTER button to save the value.
22. Press the red Back button to exit the NEW CHANNEL setup and
return to the SENSOR PROFILE menu.
23. Using the red down arrow key followed by the red ENTER button
select SAVE CHANGES to save and create current sensor profile.
24. Press the red Back button to exit the SENSOR PROFILE menu and
return to the IHH500 main menu.
25. Press the red Back button to exit the IHH500 menu and return to the
normal operation screen.
26. The normal operating screen will now be displayed with the current
peak reading, sensor output, valley reading, selected channel, gross
or tare mode, and samples per second (sps).
a.
Press the Display button to display additional information from
encoder when available.
Drawing Number: SP1187
Copyright © FUTEK Advanced Sensor Technology, Inc. Neither the whole nor any
part of the information contained in, or the product described in this manual, may
be adapted or reproduced in any material or electronic form without the prior
written consent of the copyright holder.
This product and its documentation are supplied on an as-is basis and no warranty
as to their suitability for any particular purpose is either made or implied.
This document provides preliminary information that may be subject to change
without notice.
10 Thomas, Irvine, CA 92618 USA
Tel: (949) 465-0900
Fax: (949) 465-0905
www.futek.com