background image

 

Page 

 10                                                                                                                                                            

                                                            

 

3.3

 

K

EY FUNCTIONS

 

            

     Commissioning  

The data loggers are delivered with batteries in removed condition. Insert the batteries 
to  start  the  data  logger  as  directed  in  quick  reference  manual.  Once  battery  inserted, 
Status of 

HBT

 LED starts flickering every 0.5 sec and display will come up. 

 

      

Min/Max & Tag Key 

 

 

Press this key and if the display is in OFF mode, it will start immediately, and display 

will start showing “

Min

”, “

Max

”, and current data

 

 

Keep  the  key  pressed  for  about  4  seconds  while  display  is  showing  Min  and  Max 
values, the data logger will register (log) the data (

TAG

) at that instance of time if 

batch is running 

 

Display will Show “

tAg

 

 

Continue to press the key for more than 8 seconds, and if batch is running, batch will 

stop, and display will show “

StP

. This will also stop recording further data.  

 

      

Display Sensor Type  

 

 

 

Press this key and if the display is in OFF mode, it will start immediately, and display 
will show current 

Sensor Type

 in device. 

 

After showing sensor type display will be OFF and if display in ON mode then it will 
start showing readings. 

 

Following table show display string for different sensor type. 

 

Table 4 Display strings for various sensor types

 

 
 
 
 
 
 
 
 
 
 
 
 

Sensor Type 

Display 
String 

RTD (Pt-100)

 

rtd 

4-20mA

 

4MA 

0-20mA 

0MA 

0-1V 

U1v 

J type TC 

tcJ 

K type TC 

tck 

R type TC 

tcr 

S type TC 

tcS 

T type TC 

tct 

Summary of Contents for LM-U PRO Series

Page 1: ......

Page 2: ...USE 5 2 3 TECHNICAL DATA 6 3 PRODUCT DESCRIPTION 9 3 1 STATUS LEDS 9 3 2 DISPLAY LCD 9 3 3 KEY FUNCTIONS 10 3 4 SENSOR CONNECTIONS 11 3 5 IMPORTANT INFORMATION AND GLOSSARY OF TERMS 12 4 USING THE PR...

Page 3: ...Figure 5 Batch Configuration of the Data logger 12 Figure 6 Current Temperature Reading of the LM U Pro Data logger 15 Figure 7 Downloaded data Summary 16 List of Tables Table 1 Specification 6 Table...

Page 4: ...use the product if there are signs of damage to the housing Carry out only the maintenance and repair work on this instrument that is described in the documentation Follow the prescribed steps exactly...

Page 5: ...ly and the device be examined If the device is found damaged it should not be operated and the local representative contacted for instructions Ensure that all accessories and documentation is removed...

Page 6: ...subject to cold chain requirements User can fully configure parameters like Start delay alarm high low limit sensor type temperature unit store interval etc User can also set the information like use...

Page 7: ...Alarm high low Condition Reading Types Actual with Min Max ON Display and PC Analog Input Details No of Inputs 1 Sensor Input Type Types J K R S T RTD PT 100 DIN43760 0 00385 Vdc 0 1V mA 4 20mA 0 20mA...

Page 8: ...provides information on the remaining lifetime Battery Reverse Polarity Protected General Restore Calibration User can revert to original factory calibration any time Reusability Yes Online Feature Po...

Page 9: ...Page 8 Sensor Range RTD Pt 100 100 to 600 C 4 20mA 1999 0 20mA 1999 0 1V 1999 J type TC 0 to 400 C K type TC 0 to 1200 C R type TC 0 to 1700 C S type TC 0 to 1700 C T type TC 100 to 400 C...

Page 10: ...er limit value exceeded Note For Technical reasons the display intensity of liquid crystal display becomes lower at temperatures below 0 C This has no influence on the measuring accuracy For technical...

Page 11: ...og the data TAG at that instance of time if batch is running Display will Show tAg Continue to press the key for more than 8 seconds and if batch is running batch will stop and display will show StP T...

Page 12: ...sor Connection Table Sensor Input 1 2 3 4 RTD Red White White Common White Common Volt Not connected Not connected mV Not connected Not connected 0 20 4 20 mA Not connected Not connected Note 1 Verify...

Page 13: ...set point values for the individual alarm high low trigger 4 Unit of Temperature The unit for temperature is C 5 Sensor Type User can select the required sensor input refer the table 4 from the optio...

Page 14: ...s current date and time of the data logger Click on the clock button to sync the current date and time of the device with PC time as per selected time zone 13 Start Time It is the time at which Batch...

Page 15: ...running Rec Mode user can log special event with time stamp by pressing MIN MAX TAG key for more than 4 seconds Display will show tAg Each TAG is considered as one record User can place as many TAGS a...

Page 16: ...t data report Connect the data logger to Windows PC via the Micro USB data cable The current temperature reading as per the selected sensor will be seen as shown in the figure 6 Figure 6 Current Tempe...

Page 17: ...Page 16 Figure 7 Downloaded data Summary...

Page 18: ...ack of the data logger Take the spent batteries out of the battery compartment Insert new batteries see technical data for type required into the instrument so that the positive Negative pole is visib...

Page 19: ...accordance with your local regulations Do not expose the Data Logger to extreme temperatures as it may lead to the destruction of the battery and may cause injuries 5 3 CLEANING THE INSTRUMENT If the...

Page 20: ...selection during batch configuration Sensor wire broken Display show udr If Sensor input is less than specified range which is given in Table 1 display can show udr Under the range Sensor is not inser...

Page 21: ...7 ORDER CODE Table 7 Order Code Data Logger Sensor Type Pressure Rang Communication Type Software Type Memory Type 6 6X Pro 0 Universal Input 0 No Pressure 3 USB 6 LM View Non Secure 1 Memory 7 LM vie...

Reviews: