E006
Page 5
2
S
PECIFICATIONS
2.1. Use
LM-XS-Pro E006 Data logger
is an external sensor data logger meeting the requirements of
WHO
PQS E006/TR06.3.
It stores the data up to 120 days and user can see the history data up to last
30 days on display without downloading or connecting the device to the computer. All the
parameters and alarm limits are pre-configured as per the requirement of guidelines. They have
been specifically designed for monitoring the temperature during transportation, storage of
vaccines and other medical products or the medical refrigerator products subject to cold chain
requirements.
The temperature readings are monitored and saved throughout the entire duration of
measurement program.
LMView-XS-E006 software needs to be installed in PC To download data to generate a report in
pdf format and to export data in csv format for further use.
2.2 Technical Data
Table 1 Technical Specifications
Model
LM-XS PRO E006
General
Remote Sensor
Thermistor - 10K NTC; 3mm diameter, 2.5 meter-long Cable in a sealed
cap.
Temperature
Measurement Range
-40 °C to + 80 °C (-40 °F to +176 °F): External Sensor (Fixed installation)
Accuracy
± 0.5 °C for the range -30 °C to + 30 °C
± 0.7 °C otherwise
Resolution
0.1 °C display and
storage
Unit of Measurement
Data in °C. User has an option to view the data in °F
Calibration
Each device accompanies NABL (ISO/IEC 17025) traceable certificate
Alarm
Audio-Visual.
Alarm Low Settings*
<= -0.5 °C for more than 60 minutes
Alarm High Settings*
>= 8.0 °C for more than 10 hours
Response Time
<10 min for step change of 50 °C T90 as per EN12830:1999
Logging Interval*
Measurement interval 1 minute and
Data store interval 5 minutes, pre-Fixed.
Delayed Start Option
Yes. 10 mins after start of the device
Power Requirement